VISMO monitors, Life Scope BSM-2300 series (BSM-2301, BSM-2303,
BSM-2351, BSM-2353), PVM-2701, PVM-2703 MULTI sockets and Smart cables.
In
this knowledge-sharing review, we look at the Vismo PVM-2701 and Vismo
PVM-2703 multi-parameter monitors and how they make use of the yellow
MULTI sockets. Reference is also made to the earlier Life Scope
BSM-2300K series monitors.
In
this article we reviewed the launch of the Vismo PVM-2701, and later
the PVM-2703 with a yellow sharing socket. The same yellow sharing socket accepts non-continuous semi-quantitative CO2
monitoring and is our alert for such an approach.
The first Vismo monitor was the PVM-2701 model announced by Signal 718 dated November 2009. Notice this was the first model to do away with the suffix. Vismo came from Vital Signs Monitor.
In a pilot move, the new Vismo series was designed in Japan with as many parts sourced in China. Although initial shipment were assembled and shipped from Japan, this arrangement was changed from release of PVM-2703.
This is a monitor that is short of one connector socket |
---|
The 7-parameter Vismo PVM-2703 was released in February 2011 equipped with a yellow MULTI-parameter socket for either Invasive Blood Pressure or mainstream etCO2 monitoring use. There is shortage of one connector socket, if the user could have one dedicated socket for Invasive Blood pressure and another dedicated socket for mainstream etCO2, that would be a far superior design since the monitor can now do all 7 parameters at once.
From the release of PVM-2703, shipment for both PVM-2701 and PVM-2703 models started to be directly from Shanghai, China.
The prominent feature of the VISMO PVM-2703 bedside monitor is the utilization of one yellow MULTI-parameter socket. The yellow socket does not accept ordinary measurement cables but only cables embedded with codes defined by NIHON KOHDEN.
NIHON KOHDEN had identified five types of internally configured hardware that can be linked to the MULTI-parameter sockets and to make use of these hardware, a cable with the correct code must be plugged into one of the MULTI-parameter sockets. These coded cables are collectively cited as Smart Cables by the manufacturer and the code is also known as the parameter code. Each socket selects only one channel of the hardware, except for Temperature allowing two channels of hardware to be selected.
(Of the five types of hardware configured for use by Smart Cables, only one channel of IBP hardware is configured in the VISMO PVM-2703 bedside monitor.)