NIHON KOHDEN CardioLife Acti-Biphasic Defibrillation Output Waveform

Category: Product Review, Polarity of "ActiBiphasic" waveform.

 
 
From advertising material, as well as operating and service manuals of CardioLife Acti-Biphasic defibrillators, we know the declared Acti-Biphasic waveform looks like what is shown in below picture. This is also the waveform submitted to obtain regulatory approvals around the world, supported by a Certificate of Free Sales from Japanese regulatory authority, MHLW and/or CE Certification.


The ACTI-BIPHASIC discharge output declared by the manufacturer
 
Declared characteristics of the Acti-Biphasic discharge waveform in advertising, operator and service manuals


The basic concept of a bi-phasic shock energy is to add a negative follow-up phase to the conventional mono-phasic shock to achieve the same defibrillation result using lesser energy
  
 
 
 
The characteristics of NIHON KOHDEN Acti-Biphasic defibrillation can be summarized as:
 
1. First (Launch) Phase
The Acti-Biphasic waveform is seen as operating in an open loop during the first phase (period).
 
- It is a positive truncated exponential pulse with a duration depending on the patient impedance.
- It is of a longer duration than the second period.
 
2. Second (Follow-up) Phase
The Acti-Biphasic waveform operates in a closed loop during the second phase. In a closed loop the duration of the width can be deliberately set to a specific duration.
 
- This second phase (period) is shorter in duration and of negative polarity.
- The duration in this phase is fixed to a constant 3.4ms duration regardless of patient impedance. How was
the optimal duration of 3.4ms arrived at? It is also not clear how was the maximum energy of 270 Joules concluded?
Note the first phase (period) is positive and a wider pulse than the second phase


The first to use Acti-Biphasic waveform were the CardioLife TEC-7700K series defibrillators

The CardioLife TEC-7000 series defibrillators had two models, the CardioLife TEC-7721K and CardioLife TEC-7731K (built-in pacing unit).
 
The first Acti-Biphasic defibrillators
 

Output discharged by CardioLife TEC-7700K series defibrillators is consistent with the declared waveform
 
The discharge waveform of the CardioLife TEC-7700K series on a strip chart is as shown below. The recording correctly reflects the first phase as a positive pulse.

The recording shows the voltage first swings to the top (positive saturation), then to negative saturation after some time; this is fully consistent with the official description of the Acti-Biphasic waveform.
 
The reason we are not seeing the full shape of the Acti-Biphasic waveform on the recording is because the sensitivity is set to see the smaller ECG waveform, and not the defibrillation shock which are much higher in magnitude.
 
This strip chart recording shows output of CardioLife TEC-7700K series is consistent with the declared description
 

How can we know if the Acti-Biphasic defibrillation shock actually works on patients?
The margin of error is high for data from a small 75 investigated cases
 
There is no white paper available. The Acti-Biphasic defibrillators were hurriedly launched (for export) before completion of proper clinical validation and the small sample size of seventy five investigated cases meant a high margin of error; we cannot be sure the Acti-Biphasic defibrillation shock works on patients! In addition, how can we be absolutely sure the investigated cases were done in an acceptable manner?

As the mono-phasic defibrillators are not predicate devices so the FDA 510(K) process cannot be used to clear the product for sales in the US market. Since the clinical data and methodology adopted by NIHON KOHDEN does not follow US FDA guide for safety and effectiveness, it was not possible for the Acti-Biphasic defibrillation shock to get the necessary approval for sales in the US market.
 
Nihon Kohden could have engaged a consultant to ensure a proper and acceptable clinical validation process if they were determined in obtaining US FDA approval, but the company gave up the US market.
 
The persistent remarks we often heard on American Heart Association recommendations turned out to be meaningless to Acti-Biphasic defibrillators.
 
Note current CE certification does not guarantee clinical performance since it is not included in evaluation. This means you need clinical papers in addition to CE certification, which is for safety. What is the use of safety without assurance of treatment effectiveness? It is so unfair to the patients needing immediate treatment in a life-threatening situation.
 
There are therefore two very important questions:
1. What was the hidden factor that gave the manufacturer the confidence that no further validation was needed for the Acti-Biphasic defibrillation energy?

2. Why should the manufacturer willingly give up the big US bi-phasic defibrillator market when they were already in the US mono-phasic market?


NIHON KOHDEN development team had got it all wrong

In May 1997 Nihon Kohden released the first CardioLife defibrillators with a semi-automatic AED mode for export, they were the predecessors of the AED-2100K series, discharging the same monophasic defibrillation as CardioLife TEC-7500 series defibrillators.
 
The summary of this TEC-2200K series defibrillators can be found in the 1997 Product Guide.

The monophasic CardioLife TEC-2200K series was launched for export in May 1997
 
This was a monophasic model using the non-proprietary Edmark single-phase pulse as illustrated below. This was in response to the emerging demand for AED but the use of rechargeable battery for energy made it inconvenient for public use.
 
Edwark, Single Phase Pulse defibrillation waveform used by monophasic CardioLife TEC-2200K series in 1997

 
The monophasic output discharge waveform shape is shown below.
 
CardioLife TEC-2200K mono-phasic output waveform recorded on a strip chart dated April 1995
 
 
Tokyo was unaware biphasic defibrillation was already taking hold worldwide in 1997.
 
Just a few months after the TEC-2200 series was released, then market leader Hewlett Packard made announcement to acquire Heartstream Inc. in a stock-swap deal.
 
Heartstream ForeRunner

Biphasic defibrillation waveform was becoming the new preference as it allowed for a smaller and lighter defibrillator design; more importantly it uses less current and this means less damage defibrillation will do to the heart and skin. The deal with Heartstream was how the then defibrillator market leader HP acquired biphasic technology, since to develop one would take time and speed to market was a top priority; HP was showing the way what Nihon Kohden should also do. 
 
With such a major change in the international market, there was therefore zero interest in the monophasic TEC-2200 series defibrillators offered by Nihon Kohden and the products had to be withdrawn from exporting.
 
 

 Global Demand for Biphasic Technology caught Nihon Kohden by surprise
 
It was a market disruption that NIHON KOHDEN was unprepared for. The company was at a loss for the next four years how to obtain the technology to offer biphasic defibrillators; this meant the company's defibrillator development team was helplessly doing nothing. It was obvious to any observer they could only solve the problem by buying the technology from someone.

When the demand for biphasic AEDs eventually emerged in the Japanese domestic market, the company resorted to find a suitable partner with biphasic technology for co-operation. A strategic OEM distribution agreement was announced in January 2002 that Nihon Kohden would market Cardiac Science's line of AEDs under Nihon Kohden's trade name. This arrangement was a big success and many AED-9200 and AED-9231 were sold in Japan as reflected in annual reports and presentations.
 
Mark the date January 2002, and subsequent release of first Acti-Biphasic defibrillators (CardioLife TEC-7700 series) by NIHON KOHDEN in November 2002.
  
The Cardiac Science STAR biphasic waveform was validated in accordance with US FDA guides for Safety and Effectiveness

The Acti-Biphasic defibrillation shock waveform was done in a hurry
 
To our surprise, instead of licensing the proprietary biphasic defibrillation design from Cardiac Science, a few engineers in NIHON KOHDEN could suddenly conclude a workable, proprietary Acti-Biphasic shock waveform just by playing with biphasic circuitry on pigs with minimal clinical supervision and collaboration, it is no wonder that the company had great difficulty securing the necessary clinical support to advance the number of investigated cases for proper clinical validation.
 
To date, there is not a single clinical paper published on Acti-Biphasic defibrillation.


Unsuspecting Customers were sold Unproven Defibrillators
 
Before completion of proper clinical validations, the company was bold enough to go ahead with exporting the unproven TEC-7700 series Acti-Biphasic defibrillators from November 2002, taking advantage of its established distribution network for mono-phasic defibrillators. The customers were sold unproven defibrillators by taking advantage of their good faith, because there was no published clinical paper to show that it works. This November 2002 export launch was three long years ahead of the date Japan MHLW officially approved its use for the domestic market.
 
The desperate action was taken in response to the rapid changing preference for biphasic defibrillators in the market but the process totally overlooked the seriousness of mandatory successful clinical studies before marketing; the fact that Ministry of Health, Labour and Welfare (MHLW) had not yet approved the sales of TEC-7700 series defibrillators in Japan domestic market reflected the disturbing absence of internal safeguards in corporate conduct.
 
Up to this point, the company had never exported a new product before first launching it in Japan, showing the company was in complete disarray. It is not just loss of credibility in overseas markets as a leading defibrillator exporter from Japan but a ticking time bomb with important issues left unattended.
 
Acti-Biphasic defibrillators started to be exported from 2002
 
 

  The company only started selling Acti-Biphasic defibrillators in Japan from December 2005
 
After a long period of three years for Japanese Regulatory Authority approval, NIHON KOHDEN was finally able to announce the launch of TEC-7700 series defibrillators for sales in Japan market on December 1st, 2005.
 
 

    CardioLife TEC-7700K series defibrillators were the last to use the Acti-Biphasic discharge output for export
 
CardioLife TEC-5500 series defibrillators were quickly approved for sales in Japan after the approval was given for CardioLife TEC-7700 series, based on the principle of declared substantial equivalence with the newly-approved TEC-7700 series. In Japan, the Acti-biphasic waveform discharged by Japanese version TEC-5500 series defibrillators is unlikely to be different from that discharged by TEC-7700 series defibrillators.

However, the export models of the TEC-5500 series, TEC-5600 series and TEC-8300 series defibrillators were all found to have their discharged waveform inverted, but submission documents to foreign regulatory authorities were all based on the TEC-7700 series. Are these regulatory approvals valid?
 
The "Acti-Biphasic" TEC-5500 series defibrillators started to be exported from August 2004, more than a year before the TEC-7700K series defibrillators were allowed for sales in Japan.
 
The shape of the discharge envelope is different from the TEC-7700K series defibrillators. It is a version that is flipped vertically upside down from that of the TEC-7700K series defibrillators! This was not officially announced at the time of launch.
 
Unlike the TEC-7700K series defibrillators, there was not the slightest clinical evidence to support its use. 

CardioLife TEC-5500K series started to be exported from August 2004


The Timeline

Export of CardioLife TEC-5500K series started in August 2004, more than one year before Japan MHLW actually approved the TEC-7700 series


A top prestigious University Hospital in Taiwan was the first to find the polarity of TEC-5500K discharge waveform inverted from what was declared in the operator and service manuals
  
In the next image below, we received an adverse report from a competent Biomedical Engineering Team in National Taiwan University Hospital (Taipei City) that the polarity of measured waveform discharged by two tested CardioLife TEC-5500K series defibrillators were inverted (i.e. opposite in polarity) from what the manuals had described.

There was no doubt since they had tested both models TEC-5521K (S/N 09xx4) and TEC-5531K (S/N 05xx4) to arrive at the same conclusion; the suffix K is for export models using English language as interface (for example the suffix J is for Japan domestic models), indicating more than 9000 units of TEC-5521K and more than 5000 units of TEC-5531K had been produced before the two tested units respectively. Detailed comparison was also done with defibrillators from another manufacturer (Philips) using the same testing equipment (Fluke Impulse 7000DP with 7010 Selectable Load) and the polarity was consistent with the manual descriptions of Philips.
 
This was an input from professionals that the Acti-Biphasic waveform discharged by CardioLife TEC-5500K series defibrillators starts with a negative polarity and ends with a positive polarity; it is the exact opposite polarity (inversion) of what were shown on the operator and service manuals. As far as we know, there is no known manufacturer with a discharged biphasic waveform that starts with a negative polarity, NIHON KOHDEN is unique in this approach but there is no clinical research done to validate its use on patients!
 
Is this equivalent to a delayed fixed-width mono-phasic discharge?

 

An Indirect Confirmation

The following two images showed another distributor (Thailand) sending in a Nihon Kohden defibrillator analyzer AX-103VK (OEM device) for repair.
 
The AX-103VK defibrillator analyzer has a discharge waveform output on the rear panel for oscilloscope display

The analyzer was concluded by their technical staff to be defective because the display on the oscilloscope was inverted; the analyzer was of course working fine. Said Thailand distributor is a top distributor who had sold the highest number of CardioLife TEC-7700K series defibrillators in the world and knew too well what is the "Correct Graph", confident of the defect conclusion. The conclusion turned out to be erroneous because the service manual wrongly informed them a TEC-5500K series defibrillator has similar output as a TEC-7700K series defibrillator.

Guess what? Someone (engineer) from Tokyo Support has the audacity to ask the distributor staff to "just flip the APEX/ STERNUM connections" to obtain the "correct graph"!
 
It was desperate advice, and an indirect confirmation that the discharge output waveform had been flipped intentionally.
 
Why should changing an "Evaluation machine" from a TEC-7700K defibrillator to a TEC-5500K defibrillator make a difference?

 
One likely reason could be the shape of the Acti-Biphasic waveform may have violated a certain company's rights; does the inverted waveform only apply to export models since Japan MHLW solely approved the version adopted by the CardioLife TEC-7700 series?
 
The declared current flow direction of NIHON KOHDEN Acti-Biphasic shock energy is the one on the left while we discovered actual biphasic flow is the one shown on the right

The clinical trial data cited to regulatory authorities is based on the TEC-7700 series defibrillators for all Acti-Biphasic defibrillators, including the TEC-5500 series, TEC-8300 series and latest TEC-5600 series.
 
Clinical data from TEC-7700 series defibrillators provided as support
 

It means there is no approval from regulatory authorities to use a discharge waveform that is flipped upside down, and this is a ticking time bomb.
 
Change in current direction demands fresh clinical trial and validation

As a responsible company, NIHON KOHDEN should have by now long recalled all Acti-Biphasic defibrillators from the market.
 
 
How can anyone view the discharge output of a NIHON KOHDEN defibrillator?

The QP-551VK defibrillator report viewer software can be used on a PC or laptop to review event recorded by any CardioLife defibrillator, including the AED defibrillators.
 
The defibrillator report viewer is for reviewing events recorded by a NIHON KOHDEN defibrillator

 

NIHON KOHDEN AEDs are not using the Acti-Biphasic discharge output

The first biphasic AED was exported by NIHON KOHDEN from October 2009 (Signal SE.C-19), and the model was known as AED-2100K. The new model retained many features and appearance of the Cardiac Science OEM models but claimed to discharge the same Acti-Biphasic output energy as the CardioLife TEC-7700K series defibrillators.
 
We found the two have dissimilar discharge output, and you can see a direct comparison of the two in this below image.
 

 
The AED-2100K is actually discharging a monophasic output energy similar to TEC-2200 series AED defibrillator. As we had mentioned earlier, the latter was released in May 1997, at a time when Nihon Kohden had not released any bi-phasic defibrillator.
 
Thus, like the CardioLife TEC-5500K series defibrillators, the AED-2100K defibrillator was also seen to avoid the declared Acti-Biphasic discharge waveform but in a different way.
  

CardioLife AED-2100K is exported to many parts of the world, albeit in small quantity; the quantity sold to Taiwan and South Korea is, however, substantial. It cannot be sold in the US market, similar to the CardioLife TEC-7700K series defibrillators.

CardioLife AED-2100K seen well distributed in Taiwan

The model AED-2100K did not have a screen display, which a later model was added in January 2012 with this feature. A more compact and lower-cost AED-3100 has now replaced AED-2100 series with some small changes. This means the AED-3100 is obviously not a bi-phasic AED.
 
 

The discharge output of AED-2100 is the same as older monophasic defibrillators
 
Below shows a review screen of an AED-2100 defibrillation event using the QP-551VK software; the event was recorded using a simulator with TTR value of 49 ohm. Notice the discharge output of AED-2100 is the same as older TEC-2200K mono-phasic defibrillator.
 
Where is the negative second phase?

 
For your reference, the CardioLife TEC-2200K series and discharge output waveform are shown again in below picture.
 
The TEC-2200K series with its mono-phasic discharge output waveform dated April 1995, which of course did not have a negative second phase
 
 
American Heart Association (AHA) actually has specific recommendation for mono-phasic energy shock

In the TEC-2200K series manual mode, the monophasic defibrillator energy storage has the following choices:
25J, 50J, 100J, 150J, 200J, 300J or 360J

When arrhythmia requiring defibrillation is detected in the semi-automatic mode, the defibrillator will automatically charge to the energy according to the following defibrillation sequence:
First  Defibrillation: 200J
Second Defibrillation: 300J
Third  Defibrillation: 360J
 
This is consistent with American Heart Association recommendation for a monophasic defibrillator, which Nihon Kohden is allowed to sell in the US market.

The above tells us AED-2100K, AED-2150K. AED-2151K, AED-3100 do not comply with AHA recommendations for a mono-phasic defibrillator.

Why are the operator/ service manuals not reflecting the shape we had observed?

Discharged waveform expected from AED-2100K shown on user manual
 

We should seriously pause and think about the the points raised here on the "Acti-Biphasic" defibrillation output.